DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-55 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the unit and found the x4-external helium tank prs.(psig) is broken which confirmed the reported issue.The fse has not completed the repair, thus a supplemental report will be submitted upon receipt of additional information.The full name of the event site is (b)(6).The full name of the initial reporter is (b)(6).
|
|
Event Description
|
It was reported that during a routine check, the cardiosave intra-aortic balloon pump (iabp) unit has the wrong helium indicator.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate the issue.The fse stated that an earlier inspection revealed that the helium gas status on the display was incorrect.The fse noticed that the hi press 3500 psig had a problem and was causing the helium reading to come out incorrectly, so, in order to fix the issue, the fse replaced the hi press 3500 psig (0682-00-0092-01).After the replacement, the fse found the helium indicator to be normal.The machine was then cleared for use.
|
|
Event Description
|
N/a.
|
|