Based on a preliminary review of the device log file a replacement of the cf (compact flash) card and the pba therapy control unit was recommended.Both components were returned for investigation.The cf card was inserted to a lab device whereby the reported symptom could be reproduced.The precautionary replaced pba therapy control unit was also checked within a long-lasting test but did not reveal any abnormalities.If the pba therapy control unit no longer can access the cf card the device performs a reboot.The therapy is discontinued for a maximum of 16 seconds, before therapy will be resumed with the last valid settings.If the reboot is not successful an acoustical alarm will be given and a "system failure" message is posted as it could be confirmed for the case in question.Manual ventilation remains possible.Finally, it can be concluded that a faulty cf card was root cause of the reported symptom.The device has reacted as specified by posting corresponding alarms to inform the user about the situation.The faulty cf card was replaced and the device returned to use without further problems reported.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|