• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. M2A-T UNIV 2-HOLE SHL 50MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. M2A-T UNIV 2-HOLE SHL 50MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2015
Event Type  Injury  
Event Description
It was reported patient underwent a revision surgery approximately 9 years post implantation due to unknown reasons.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).Concomitant medical devices: 15-105011- m2a tpr hi carbon 37/28mm lnr- 810970; 11-103200- taperloc por lat fmrl 5.0x130- 042600; 103532- ti low profile screw 6.5x25mm- 286550; 103532- ti low profile screw 6.5x25mm- 317410.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 00727, 0001825034 - 2022 - 00730, 0001825034 - 2022 - 00731, 0001825034 - 2022 - 00732.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product location unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Device location unknown.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-T UNIV 2-HOLE SHL 50MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13964701
MDR Text Key288282788
Report Number0001825034-2022-00728
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Model NumberN/A
Device Catalogue Number15-103650
Device Lot Number232620
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-