• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25REC
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2022
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during other non clinical activity, they noticed scratches / flow lines on the reservoir.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on mar 31, 2022.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 11, 3331, 4114, 3221, 4315).Type of investigation #1: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #2: 3331 - analysis of production records.Type of investigation #3: 4114 - device not returned.Investigation finding: 3221 - no findings available.Investigation conclusions: 4315 - cause not established.The affected samples were not returned for evaluation.Previous complaints were reviewed with their samples received and confirmed the flow lines were present; however, once measured, they were found to be within specification.Representative retention samples were inspected and noted to have flow lines that were measured and found to be within specification.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key13966901
MDR Text Key289199216
Report Number1124841-2022-00076
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450820
UDI-Public(01)00699753450820
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberZG14
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-