MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Infusion or Flow Problem (2964); Insufficient Information (3190)
Patient Problems Chills (2191); Diaphoresis (2452)
Event Date 01/01/2022
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: (b)(6) 2022, explanted: (b)(6) 2022 product type catheter.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 07-oct-2023, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacturer representative regarding a patient who was receiving morphine (10mg/ml a t 2.399mg/day) and clonidine (765mcg/ml at 183.52mcg/day) via an implantable pump for non-malignant pain and spinal pain indications.It was reported that the patient stated they were playing golf approximately 3 weeks ago and afterwards they had a return of pain, sweats and chills.The healthcare provider (hcp) accessed the catheter access port (cap) with no flow approximately 2-3 weeks ago.The cause of the event was undetermined.The patient had a catheter revision on (b)(6) 2022.During surgery, the hcp gently pulled on the spinal segment which reportedly had some resistance, until it became easier to access with no resistance.The spinal segment was explanted and replaced with a new 8782.The hcp then aspirated 2ml from cap and the event was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13970822
MDR Text Key290110114
Report Number3004209178-2022-04093
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00763000422608
UDI-Public00763000422608
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2023
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Date Manufacturer Received03/28/2022
Date Device Manufactured11/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexMale
Patient Weight94 KG
-
-