Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
|
|
Event Description
|
On (b)(6) 2022, getinge became aware of an issue with one of our surgical lights ¿ hanaulux 2003.As it was stated, the headlight fell on ground.No injury has been reported due to described issue, however, we decided to report this case based on potential as any part falling down could lead to serious injury.
|
|
Event Description
|
On (b)(6) 2022, getinge became aware of an issue with one of our surgical lights ¿ hanaulux 2004.As it was stated, the locking ring loosened and went missing, resulting in detachment of the examination light (headlight, spring arm and extension arm).No injury has been reported due to described issue, however, we decided to report this case based on potential as any part falling down could lead to serious injury.
|
|
Manufacturer Narrative
|
According to the reporting timeframe we would like to provide the information about current status of the issue.Please be advised that it is being investigated.Additional information will be provided following the conclusion of the investigation.The correction of d1 brand name and b5 describe event or problem fields deems required.This is based on the additional information that has been received.Previous b5 describe event or problem: on 30th of march 202, getinge became aware of an issue with one of our surgical lights ¿ hanaulux 2003.As it was stated, the headlight fell on ground.No injury has been reported due to described issue, however, we decided to report this case based on potential as any part falling down could lead to serious injury.Corrected b5 describe event or problem: on 30th of march 2022, getinge became aware of an issue with one of our surgical lights ¿ hanaulux 2004.As it was stated, the locking ring loosened and went missing, resulting in detachment of the examination light (headlight, spring arm and extension arm).No injury has been reported due to described issue, however, we decided to report this case based on potential as any part falling down could lead to serious injury.Previous d1 brand name: hanaulux 2003.Corrected d1 brand name: hanaulux 2004.
|
|
Search Alerts/Recalls
|
|
|