Model Number 186715000
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/04/2022
Event Type  Injury  
Manufacturer Narrative
Additional narrative: additional device product codes: kwp; kwq; mnh; mni; osh. Complainant part is not expected to be returned for manufacturer review/investigation. Initial reporter is a j&j sales representative. Without a lot number the device history records review could not be completed. Product was not returned. Based on the information available, it has been determined that no corrective and/or preventative action is proposed. This complaint will be accounted for and monitored via post market surveillance activities. If additional information is made available, the investigation will be updated as applicable. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2022 during spinal fusion procedure set screw stripped and potentially left metal shards in the patient. Entire screw and set screw were replaced. Surgery was completed with a delay of thirty (30) minutes. Patient outcome is unknown. This report is for one (1) mis single inner setscw. This is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMIS SINGLE INNER SETSCW
Type of DeviceORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ 02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ 02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key13978037
MDR Text Key288389159
Report Number1526439-2022-00518
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034121192
UDI-Public(01)10705034121192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number186715000
Device Catalogue Number186715000
Device Lot Number322413
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2021
Is the Device Single Use? Yes
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 04/01/2022 Patient Sequence Number: 1
Treatment
VIPER UNI SCREW 6X40MM TI; VIPER UNI SCREW 6X40MM TI; WIDE BLADE HOOK
-
-