• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS Back to Search Results
Model Number 380677-17
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2022
Event Type  malfunction  
Manufacturer Narrative
An intuitive surgical, inc. (isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint. The reported complaint was not confirmed based on the field evaluation. The fse cleaned all blue fiber connections, pulled logs and test drove the system and did not find any error codes that could interfere with intuitive motion. The fse tested and verified that the surgeon side console (ssc) had intuitive motion and confirmed that the system was ready for use. The fse performed service to the system, but no part replacement was required. A review of the site's complaint history does not reveal any additional records involving this product and/or this event. No image or video clip for the reported event was submitted for review. A review of the site's system logs for the reported procedure date was conducted by isi technical support engineer (tse) and field service engineer (fse). Investigation revealed no related errors to the reported issue. This complaint is being classified as a reportable event due to the following conclusion: the customer converted to the secondary surgeon side console (ssc) after the start of the procedure due to a lag on the ssc when using the master tool manipulator or when controlling a specific universal surgical manipulator. While there was no harm or injury to the patient, system unavailability after the start of a surgical procedure could likely cause or contribute to an adverse event if it were to recur as it may lead to an injury due to the patient¿s inability to tolerate a conversion.
 
Event Description
It was reported that during a da vinci-assisted ventral hernia ipom surgical procedure, the surgeon encountered a lag on the surgeon side console (ssc) resulting in the surgeon converting from ssc 719167 to ssc 704963 which resolved the issue. The surgeon was uncertain if the lag was on both the master tool manipulators (mtms) and the universal side manipulators (usms). No troubleshooting was performed, only switching of the consoles. The technical support engineer confirmed there were no errors in the logs. The procedure was completed with no reported injury. Intuitive surgical, inc. (isi) contacted the da vinci coordinator and obtained the following additional information about the complaint: it was confirmed that the procedure was not converted, it was a dual console therefore the surgeon switched to the second console. There were no post operative complications for the patient. The customer also confirmed that the system functionality was checked when powering on. The surgeon resolved the issue by switching consoles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDAVINCI XI
Type of DeviceSURGEON SIDE CONSOLE, SMART PEDALS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key13992440
MDR Text Key298253388
Report Number2955842-2022-10875
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110744
UDI-Public(01)00886874110744
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number380677-17
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Date Manufacturer Received03/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/27/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A

Patient Treatment Data
Date Received: 04/03/2022 Patient Sequence Number: 1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-