• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 50; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 50; LAMP, SURGICAL Back to Search Results
Model Number ARD568604998
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
On (b)(6) 2022 getinge became aware of an issue with one of surgical lights - lucea 50.It was stated the cover was cracked and the particles was missing.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
 
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
 
Manufacturer Narrative
Getinge became aware of an issue with two surgical lights lucea 50.It was stated that covers were cracked and the particles were missing.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, surgical lights did not meet their specification, since cracks in covers, resulting in missing plastic particles could be considered as technical deficiency, and in this way devices contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.When reviewing reportable events for this type of issues we were able to establish that the received incident is occurring at a low ratio.We have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.According to pictures provided, plastic top covers are deteriorated in corners and a broken parts are missing.The damages were probably caused by a collision.A new covers available as spare parts in the kit must be installed in order to replace damaged covers and to avoid any incident.To prevent any safety issue the lucea 50/100 instruction for use (01741 rev.10, pages 26-27) mentions to check the light heads for chipped paint, impact marks and any other damage, during the daily check.We believe the related devices are performing correctly in the market.We also believe that if the manufacturer recommendation would have been followed the incident would have been avoided.Getinge shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 50
Type of Device
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key13993461
MDR Text Key288483497
Report Number9710055-2022-00104
Device Sequence Number1
Product Code FTD
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568604998
Device Catalogue NumberARD568604998
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/19/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/16/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-