• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SYNERGY RESECTION SHAVER CONSOLE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SYNERGY RESECTION SHAVER CONSOLE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number SYNERGY RESECTION SHAVER CONSOLE
Device Problems Overheating of Device (1437); Smoking (1585); Protective Measures Problem (3015)
Patient Problem Insufficient Information (4580)
Event Date 03/04/2022
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On 03/11/2022, it was reported by a distributor via sems that while using the ar-8305 shaver console, the handpiece started to heat almost smoking, and displayed error code c34.This was discovered during use in a procedure.Requested additional information.Additional information 03/16/2022 on 03/18/2022, it was reported by a distributor via email that while using the ar-8305 shaver console, the handpiece was heating the surgeon hand, and displayed error code c34.This was discovered during use in a rotator cuff repair procedure on (b)(6).The case was completed by only changing the burr (not the handpiece), and it solved the problem.(same lot number for the 2 burrs).
 
Manufacturer Narrative
The complaint is not confirmed.The reported device was not returned for investigation.No pictures were provided.However, per the event description is inferred that the cause for the reported failure is wear and tear: the c34 error is an indication of an overcurrent above 7.0a for 3 seconds.An overcurrent condition can be caused by a shaver handpiece motor that has reached end of life.The increased temperature on the shaver handpiece is also consistent with end of life.Per synergy and adapteur power system¿ ii (aps ii) shaver handpieces direction for use, dfu-0154-5 section f.Inspection and maintenance, the user is instructed to inspect the device for excessive heating, and the user is informed that the device may have reached end of life if critical errors or abnormal temperature are noticed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY RESECTION SHAVER CONSOLE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13995218
MDR Text Key296784106
Report Number1220246-2022-04698
Device Sequence Number1
Product Code HWE
UDI-Device Identifier00888867123854
UDI-Public00888867123854
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K932699
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSYNERGY RESECTION SHAVER CONSOLE
Device Catalogue NumberAR-8305
Device Lot Number10210943
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/18/2022
Date Device Manufactured05/17/2018
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-