• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. UNKNOWN ENT REFLEX ULTRA PLASMA WAND AND TURBINATOR WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. UNKNOWN ENT REFLEX ULTRA PLASMA WAND AND TURBINATOR WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Adhesion(s) (1695)
Event Date 01/01/2008
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case-(b)(4).Wei mengling, li lingbo, & lu liubin.(2008).Clinical observation of 93 patients who underwent modified nasal septum deviation correction.Jilin medical journal, 29(1), 49-49.
 
Event Description
It was reported that on literature review clinical observation of 93 cases of modified nasal septum deviation correction, 2 patients had adhesions between the nasal septum and inferior turbinate between 2 weeks to 1 month after a nasal septal deviation correction under nasal endoscopy procedure using reflex coblation wand.The events were treated/resolved by separating the adhesions.Patients healed after the separation.No further information is available.
 
Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A complaint history review found no similar reported events.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per case details no further information is available.Without supporting clinical/medical documents, a thorough investigation cannot be performed.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Should any additional clinical information be provided this complaint will be re-evaluated.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed, and the root cause could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ENT REFLEX ULTRA PLASMA WAND AND TURBINATOR WAND
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13995595
MDR Text Key288594682
Report Number3006524618-2022-00150
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122652
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Literature,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-