Catalog Number 511.701
Device Problems Noise, Audible (3273); Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2022
Event Type  malfunction  
Event Description
It was reported from (b)(6) that during service and evaluation, it was determined that the compact air drive device moving parts did not move smoothly, the mode switch resistance was low, made excessive nose, had low power and had an air leak.It was further determined that the device failed pretests for free movement of soft mode switch, noticeable noise, power with power test bench and air leak.It was noted in the service order that the device did no work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.An evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT AIR DRIVE II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key13999849
MDR Text Key288602418
Report Number8030965-2022-02168
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819085268
UDI-Public07611819085268
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2022
Date Manufacturer Received04/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-