OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0720
Device Problems Break (1069); Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus for evaluation.Upon inspection and testing of the device, it was confirmed that the cutting wire (knife wire) was broken.The coated portion of the cutting wire was torn, and the broken portion was scorched and melted.The outer diameter of the cutting wire was measured and no abnormalities were observed.The length of the coated portion of the cutting wire, and the cutting wire itself, presented no abnormalities.There were no missing parts in the subject device.There were no abnormalities identified that could have led to the breakage of the cutting wire.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.Based on the results of the legal manufacturer's investigation, it is likely that output was activated while the torn area of the coated portion of the cutting wire was in contact with the distal end of the endoscope while the forceps elevator was raised, causing the cutting wire to become instantly hot, resulting in breakage of the cutting wire.It is likely the coated portion of the cutting wire was torn while the forceps elevator of the endoscope was raised, the slider was pushed more than needed, causing the cutting wire to deflect.The coated portion of the cutting wire and the metal part of the distal end of the endoscope came into contact.Then, the cutting wire was moved back and forth, causing the coated portion of the cutting wire to tear.The device's instruction manual provides the following warnings which may help to prevent the issue: "¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿ when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion." olympus will continue to monitor field performance for this device.
 
Event Description
The customer reported that during a therapeutic endoscopic sphincterotomy (est) procedure, the knife wire broke when it was activated.Another similar device was used to complete the intended procedure.There was no report of patient or user injury due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Type of Device
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14000729
MDR Text Key298281647
Report Number8010047-2022-05633
Device Sequence Number1
Product Code KNS
UDI-Device Identifier04953170382628
UDI-Public04953170382628
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K950166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKD-V411M-0720
Device Lot Number1YK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2022
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ERBE VIO300D ELECTROSURGICAL SYSTEM
-
-