• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM; GAUGE, DEPTH Back to Search Results
Model Number SD319.002
Device Problems Material Deformation (2976); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date in 2022, during post op inspection it was identified that depth gauges were stiff and don't slide and the locking compression plate has stripped locking hole.There was no know patient involvement.This report is for one (1) depth gauge for 2.0mm and 2.4mm screws-70mm.This is report 4 of 5 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key14014113
MDR Text Key290159472
Report Number2939274-2022-01194
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982004925
UDI-Public(01)10886982004925
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSD319.002
Device Catalogue NumberSD319.002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-