Catalog Number 8065752451
Device Problem Fluid/Blood Leak (1250)
Patient Problem Retinal Injury (2048)
Event Date 10/01/2021
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported while indenting the globe a fresh retinal foramen was found.The foramina were treated with laser the eye was supplied with gas and patient had to camp.No hospitalization was required.The postoperative course was uncomplicated.Additional information received from the company representative indicating during the surgery all the 27g and 25g trocars were spontaneously leaking immediately after insertion leaked out of the valves.There were multiple products reported from the site and it is still unknown if this actual product is associated with the complaint.This report is third of the of three reports.This one represents the 25 gauge valved trocars.
 
Manufacturer Narrative
Additional information provided in h.6.And h.10.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.Investigations have been completed and manufacturing process enhancements have been implemented in order to improve the performance of the valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14014492
MDR Text Key288606283
Report Number1644019-2022-00253
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657524518
UDI-Public00380657524518
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Catalogue Number8065752451
Device Lot Number2362495H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED CUSTOM PAK, BEL
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-