Catalog Number 955607
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Event Description
It was reported that a broken wheel was observed on an ak 98 machine during an unspecified therapy step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was not received for evaluation; however, the device was evaluated on site by a qualified technician.Upon evaluation it was observed the base of the molding that holds the wheel was cracked and the wheel shaft was broke off.Photographs of the sample were also provided for evaluation.Visual inspection of the pictures showed the wheel disconnected from the foot unit.The reported condition was verified.The possible cause of the condition could be use related as it appeared the unit was run into something at a high-speed causing the crack and break.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDOLLA MODENA MONITORS
via modenese, 66
medolla modena 41036
IT   41036
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14015209
MDR Text Key291952730
Report Number9616240-2022-00004
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124519
UDI-Public(01)07332414124519
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201809
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/09/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-