• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number 1002717
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2022
Event Type  malfunction  
Manufacturer Narrative
No product was returned for evaluation. Should new relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the pump delivered a bolus without user input. There was no adverse impact to customer's blood glucose level. No additional patient or event information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameT:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of DeviceAUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key14017303
MDR Text Key289899493
Report Number3013756811-2022-37498
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K201214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/28/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1002717
Device Catalogue Number1004219
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-