MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-55
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2022
Event Type  malfunction  
Manufacturer Narrative
Acquiring additional information about the status of unit.A supplemental report will be submitted upon completion of our investigation.Not returned to manufacturer.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) battery will not charge and remained in hybrid mode.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
It was reported that prior to use on a patient, the cardiosave intra-aortic balloon pump (iabp) battery will not charge and remained in hybrid mode.There was no patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields; b4, g3, g6, h2, h4, h6, h10, h11.Corrected fields; h3.A getinge field service engineer (fse) was dispatched to evaluate the unit and was able to reproduce the reported issue.All tests were done with a test balloon and a simulator.The fse noticed a liquid on the mechanical part which allows the cardiosave to switch to "rescue or hybrid" mode.So the fse cleaned this part and the sensor.The unit was released to the customer for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14028565
MDR Text Key290088255
Report Number2249723-2022-00743
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-