Updated fields: b4, g4, g7, h2, h10, h11 corrected fields: d10, h3, h6 (medical device ¿ problem code, type of investigation).Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to evaluate the iabp and was unable to reproduce the reported issue.The fse ran the unit for 2 hours and found no issue.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|