AESCULAP INC. STEAM STER LOCKS ORANGE; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS Back to Search Results
Model Number US906
Device Problem Unexpected Color (4055)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2022
Event Type  malfunction  
Event Description
It was reported to aesculap inc.That a steam ster locks orange (part # us906) was used to secure a sterilization container.According to the complainant, the indicator dot transitioned to its post sterilization color, however, the indicator dot turned back to blue after approximately thrity (30) to sixty (60) days while in storage in the sterile processing department (spd).The lock was not in extreme temperatures while in sterile storage within the department.The containers are stored on normal shelving and are not exposed to intense light.The complaint device has been returned to the manufacturer for evaluation.No patient involvement.
 
Manufacturer Narrative
Manufacturer evaluation: investigation on-going.Should relevant additional / investigation results become available, a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
Manufacturer evaluation: the device was returned to the manufacturer for physical evaluation.A visual examination of the chemical indicator revealed that one (1) of the indicator dots was grey/blue and one (1) was light blue.Both dots were not consistent with the expected initial or final transition color.The device history records (dhr) were not able to be reviewed as no lot # was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.The investigation into the cause of the reported problem was able to confirm the failure mode of chemical indicator failed to transition to its post sterilization color.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEAM STER LOCKS ORANGE
Type of Device
INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
jonathan severino
3773 corporate parkway
center valley, PA 18034
4847197287
MDR Report Key14053537
MDR Text Key289016528
Report Number2916714-2022-00039
Device Sequence Number1
Product Code JOJ
UDI-Device Identifier05055049091927
UDI-Public5055049091927
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K890759
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUS906
Device Catalogue NumberUS906
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-