Model Number 3CX*FX25REC |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
|
|
Event Description
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during out of box, the caps were broken / smashed.No patient involvement.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on april 8, 2022. upon further investigation of the reported event, the following information is new and/or changed: d10 (device availability - added date returned to manufacturer); g4 (date received by manufacturer); g7 (indication that this is a follow-up report) ; h2 (follow-up due to additional information); h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information and device evaluation) ; h3 (device evaluated by manufacturer) ; h4 (device manufacture date) ; h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The affected sample was inspected upon receipt to confirm damage to the non vented cap attached the arterial pigtail line.All capiox units are 100% visually inspected at several points in the production process.A representative retention sample was inspected to confirm no damage to the unit specifically with the arterial pigtail line.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Search Alerts/Recalls
|
|