Catalog Number 383951 |
Device Problem
Retraction Problem (1536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/15/2022 |
Event Type
malfunction
|
Event Description
|
It was reported bd pegasus gn 18ga x 1.16in prn had difficult needle disengagement.The following information was provided by the initial reporter: "when the user is preparing to withdraw the needle, it is difficult to withdraw the needle core.".
|
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported bd pegasus gn 18ga x 1.16in prn had difficult needle disengagement.The following information was provided by the initial reporter: "when the user is preparing to withdraw the needle, it is difficult to withdraw the needle core.".
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 04-apr-2022.H.6.A device history review was conducted for lot number 1230464.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, a fully activated device was submitted to aid in our investigation.Unfortunately, due to the activated safety mechanism our engineers were unable to test this specific device for disengagement performance; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
|
|