This device has not been returned.However, investigation determined that the minute ventilation (mv) sensor in ingenio devices has the potential to increase the pacing rate to the maximum sensor rate more quickly than expected.As a result, a software update was released in 2013 that reduced the minimum mv response factor settings and re-distributed the remaining mv response factors to provide a consistent change in sensor rate across the range of mv response factors.This corrective action is expected to reduce, but not eliminate the occurrence of this behavior.
|
It was reported that this patient presented in the emergency room due to experiencing tachycardia of 120bpm.The pacemaker was interrogated and it was noted that the device was pacing on maximum sensor rate when the patient was lying in a lateral position.Additionally, this pacemaker recorded two signal artifact monitoring (sam) episode.It was suspected that the patients tachycardia was induced by the minute ventilation (mv) sensor due to the physician turning off the mv feature and the tachycardia stopping and when the mv feature was turned back on, the tachycardia appeared a few seconds later.The patient was sent home with the device programming mv sensor disabled.Upon an office visit a few days later, the right atrial (ra) lead exhibited low out of range pacing impedance measurements, measuring less than 200 ohms.It was suspected that there was an issue with the ra lead insulation and noise was also noted on the atrial channel.Several tests were performed and the device was reprogrammed.Boston scientific technical services reviewed the available data and provided programming options and suggested lead replacement.No additional adverse patient effects were reported.This pacemaker and unknown ra lead remains in-service.
|