Model Number U228
Device Problems Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2021
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system triggered a lead safety switch due to high out-of-range right atrial (ra) pacing impedance of greater than 2500 ohms. No capture at maximum outputs and noise on stored electrograms were also noted. All lead measurements were normal in unipolar programming and the patient has a good underlying rhythm, so the physician elected to keep the current programming. No significant pauses noted. No adverse patient effects were reported. This device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVISIONIST X4 CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14069048
MDR Text Key288999435
Report Number2124215-2022-09577
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/22/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/22/2022
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number750339
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-