Model Number ROB10024
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a cori assisted tka surgery, the real intelligence cori software was showing a 5 degree flexion contracture when surgeon felt that the knee was at 0 degrees. The plan showed loose lateral compartment when they planned for it to be 1 mm tight. This was noticed before the bone cuts were made, the surgeon cut what he saw and then recut to fix the issue (no additional unintended bone cuts were performed). The procedure was completed with the same device without delay. Patient outcome was acceptable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREAL INTELLIGENCE CORI
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14074257
MDR Text Key289022284
Report Number3010266064-2022-00279
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 04/11/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-