Model Number L107758
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2022
Event Type  malfunction  
Event Description
It was reported that the mobile power unit did not boot up.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed. .
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the mobile power unit (mpu) not booting up was confirmed.During the evaluation of the returned mpu, serial (b)(6), the reported event was verified; the system did not boot up.No functional test was performed due to the unit not powering on.The power supply pcb was replaced and preventative maintenance was performed.The damaged pcb was forwarded to product performance engineering (ppe) for further evaluation.Visual inspection of the returned mpu power supply pcb revealed an open fuse f1.The fuse was bypassed causing the circuitry to be damaged on power on.Due to the damage endured by the board, no further troubleshooting was performed.A root cause for the mpu not booting up was due to a damaged power supply pcb; however, the root cause for the damaged board was not conclusively determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or quality assurance specifications.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.Heartmate 3 patient handbook section 6 ¿caring for the equipment¿ and heartmate 3 instructions for use (ifu) section 8 ¿equipment storage and care¿ explain how to care for and clean all equipment, including the mpu.Heartmate 3 patient handbook section 10 ¿safety checklists¿ and heartmate 3 instructions for use (ifu) section e ¿safety checklists¿ provide the user with checklists to assist the patient in performing routine maintenance of all components of the heartmate 3 left ventricular assist device, including the mpu.Heartmate 3 instructions for use section 4-¿system monitor¿ and heartmate 3 patient handbook section 6-¿caring for the equipment¿ explain how to properly handle the equipment to prevent damage.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE MOBILE POWER UNIT, EU
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key14084053
MDR Text Key289095983
Report Number2916596-2022-10332
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P060040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberL107758
Device Lot Number8190666
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse