BAXTER HEALTHCARE CORPORATION INTERLINK SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C6425
Device Problems Backflow (1064); Complete Blockage (1094); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2022
Event Type  malfunction  
Event Description
It was reported that the gummy stopper (injection port) of an interlink system solution set became dislodged and went into the tubing clogging it.When this occurred, the medication backed up into the iv (intravenous) bag and fluids would not run.The issue was identified while the nurse was injecting medication into the y-port.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Device manufacturer address g1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was done which observed a recessed gland in the second y-site interlink.Functional testing including pressure and clear passage tests were performed which revealed blockage in the sample due to the recessed gland.A backflow functional test was performed with an in-lab secondary set which revealed backflow of solution in the device.The reported condition was verified.The cause of the condition was due to a manufacturing assembly issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14090803
MDR Text Key289182427
Report Number1416980-2022-01774
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412002149
UDI-Public(01)00085412002149
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C6425
Device Lot NumberR21F24069
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LACTATED RINGERS SOLUTION
Patient SexMale
-
-