MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: (b)(6) 2015, product type: catheter. Product id: 8781, serial# (b)(4), implanted: (b)(6) 2022, product type: catheter. Product id: 8781, serial# (b)(4), implanted: (b)(6) 2022, product type: catheter. Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 14-may-2017, udi#: (b)(4); product id: 8781, serial/lot #: (b)(4), ubd: 02-sep-2023, udi#: (b)(4); product id: 8781, serial/lot #: (b)(4), ubd: 20-dec-2023, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving baclofen via an implanted pump. It was reported that the hcps were questioning the patency of the catheter. A dye study was scheduled for (b)(6) 2022 but was canceled and a revision was completed on (b)(6) 2022. Catheter aspiration was attempted at the beginning of the catheter revision, and it was unsuccessful. During the revision, part of the catheter was replaced. After revising the catheter, the hcp was able to successfully aspirate the catheter assembly. There were no reported environmental, external, or patient factors that may have led or contributed to the issue, and the issue was noted to be resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14101683
MDR Text Key289175470
Report Number3004209178-2022-04649
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured06/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/13/2022 Patient Sequence Number: 1
-
-