Model Number 1002717
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2022
Event Type  malfunction  
Manufacturer Narrative
Although the user initially reported the event as occurring on (b)(6) 2022, the user later claimed the event took place on (b)(6) 2022, after the initial awareness date of the complaint. An investigation of pump data logs was performed for both dates. Tandem quality engineer evaluated pump data and concluded the following: there is no evidence that the pump experienced a malfunction or failure. No product was returned for evaluation. Should new relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the pump delivered a bolus without user input. There was no adverse impact to customer's blood glucose level. A pump log review was performed, no device malfunction or failure was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameT:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of DeviceAUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key14104182
MDR Text Key289211917
Report Number3013756811-2022-37547
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K201214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/28/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1002717
Device Catalogue Number1004219
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-