• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE Back to Search Results
Model Number M0063903200
Device Problems Peeled/Delaminated (1454); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2022
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The returned stone cone was analyzed, and a visual inspection noted that the device was returned with the coil was exposed.The green coating of the coil was peeled near the distal end.The stone cone was unable to be closed fully.No other problems were noted.The reported event was confirmed.Based on all available information, it is most likely that the damage was caused due to procedural factors such as handling, technique, device manipulation or interaction with another device.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that a stone cone was used in the kidney during an stone retrieval procedure performed on (b)(6) 2022.During the procedure, the stone cone did not open inside the patient.Another stone cone was used to complete the procedure.Investigation results revealed that the coil was peeled near the distal end; therefore, this is now an mdv reportable event.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONE CONE
Type of Device
DISLODGER, STONE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14135508
MDR Text Key289446214
Report Number3005099803-2022-02115
Device Sequence Number1
Product Code FGO
UDI-Device Identifier08714729430223
UDI-Public08714729430223
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM0063903200
Device Catalogue Number390-320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-