A review of the dhr and inspection records was conducted, and no similar concerns were found.One catheter was returned for review.Visual inspection did not find any damage to the catheter.The catheter was leak tested with a water filled syringe using red food coloring.Evidence of leakage was confirmed on the catheter tubing.Upon closer observation under magnification, a small pin hole was observed.The most probable root cause for the damage to the catheter was most likely related to an event within the user environment.Possibly the catheter came in contact with a sharp instrument.Per the event description from the customer, the hole in the catheter was observed after removing the catheter.It was not stated how long the catheter had been in place when the white spot on the baby's leg appeared.There have been no other complaints regarding this issue with this lot number.Since the reported issue was most likely related to the user environment and not a manufacturing error, no corrective action will be taken.
|