• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 10623
Device Problems Difficult to Advance (2920); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2022
Event Type  malfunction  
Manufacturer Narrative
The synergy ous mr 3.00 x 38mm stent delivery system was returned for analysis.An examination (visual and via scope) of the crimped stent found stent damage.Stent struts in the proximal region were found to be lifted from their crimped position and pulled distally.The undamaged crimped stent outer diameter was measured and the result is within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube identified multiple hypotube kinks.A visual examination of the outer and inner lumen and mid-shaft section found no issues.The device was successfully loaded onto a 0.014 inch guidewire with no issues noted.
 
Event Description
It was reported that stent damage occurred.The target lesion was located in the severely calcified diagonal and left circumflex arteries.A 3.00 x 38 synergy drug-eluting stent was introduced but had difficulty advancing over the guidewire.The device was removed and found to be kinked in the stent.The procedure was completed with a different device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14148986
MDR Text Key289693574
Report Number2134265-2022-04331
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/02/2023
Device Model Number10623
Device Catalogue Number10623
Device Lot Number0026940895
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE CATHETER: XB 3.5
-
-