• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY KIT, RADIUS-7; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY KIT, RADIUS-7; OXIMETER Back to Search Results
Model Number 24949
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/24/2022
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
The customer reported the monitor display was frozen.No consequences or impact to patient were reported.
 
Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
 
Event Description
The customer reported the monitor display was frozen.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT, RADIUS-7
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key14149344
MDR Text Key290156723
Report Number3019388613-2022-00097
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997009249
UDI-Public00843997009249
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140188
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24949
Device Catalogue Number9664
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-