MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problems Computer Software Problem (1112); No Display/Image (1183); Power Problem (3010)
Patient Problem Hyperglycemia (1905)
Event Date 04/13/2022
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the insulin pump had a insulin pump error alarm, battery failed alarm, also had blank screen.No harm requiring medical intervention was reported.The customer will continue to use the device.
 
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
The insulin pump p-cap / test reservoir locks in place properly.The pump powered up properly after battery installation.No blank display noted.The pump passed the displacement test, sleep current measurement, active current measurement and self test.The pump was monitored and no unexpected powering off or blank display noted.Successfully downloaded history files and traces using thus software.Power management tool confirmed unloaded vlith voltage and loaded vlith voltage is within specs.However, pump error 53 was confirmed in history file on 04/12/2022 19:48:16.000.Pump was cut open to perform visual inspection and found evidence of moisture damage on the keypad connector on j1/pcba1, force sensor, motor and battery tube assembly noted.The following were noted during visual inspection: scratched lcd window, missing display window cover, cracked keypad overlay, cracked case (battery tube) and broken belt clip rails in summary, customer alleged blank display, failed battery test not confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key14154336
MDR Text Key298947111
Report Number2032227-2022-181116
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316631
UDI-Public(01)000000763000316631
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG41DS0
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received07/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
Patient Weight93 KG
-
-