• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE INFUSION SET INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE INFUSION SET INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 2426-0007
Device Problem Failure to Infuse (2340)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2022
Event Type  malfunction  
Event Description
It was reported that the bd alaris pump module smartsite infusion set experienced flow issues. The following information was provided by the initial reporter: defective tubing. Attempted to start administration of arsenic iv infusion. Pump kept beeping "check tubing" and was unable to run medication. Double checked all clamps were open. Picc line flushed again - no resistance, positive blood return. Replaced the channel, still didn't work. Then, retrieved new primary tubing, and pump was then able to administer infusion.
 
Manufacturer Narrative
Device expiration date: unknown device manufacture date: unknown investigation summary: a complaint of pump alarming "check tubing" during infusion was received from the customer. No product or photo was returned by the customer. The customer complaint of tubing defective could not be verified due to the product not being returned for failure investigation. A device history record review could not be performed on model 2426-0007 because the lot number is unknown. Due to no sample being received, an investigation could not be performed, and a root cause could not be determined. This incident has been added to our database of reported incidents. Our business team regularly reviews the collected data for identification of emerging trends. Due to no sample being received, an investigation could not be performed, and a root cause could not be determined. No product will be returned per customer. No investigation was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD ALARIS PUMP MODULE SMARTSITE INFUSION SET
Type of DeviceINTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14162939
MDR Text Key289735997
Report Number9616066-2022-00448
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403227998
UDI-Public10885403227998
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number2426-0007
Device Catalogue Number2426-0007
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-