STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 0DEG 30MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU Back to Search Results
Catalog Number NLS-300000B
Device Problem Device-Device Incompatibility (2919)
Patient Problems Foreign Body Reaction (1868); Inflammation (1932); Metal Related Pathology (4530)
Event Date 03/25/2022
Event Type  Injury  
Manufacturer Narrative
Voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported elevated levels is considered to be under the scope of this recall.No further investigation is required.Device not returned to the manufacturer.
 
Event Description
Plaintiff was implanted with a left rejuvenate modular hip stem.It¿s further alleged that blood work revealed elevated levels of cobalt and chromium.The patient¿s left hip was revised.
 
Manufacturer Narrative
Reported event: an event regarding abnormal ions level involving a rejuvenate modular device was reported.The event was confirmed.Method & results: -device evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned -device history review: review of device history records found the devices in the reported lot were accepted into final stock with no reported discrepancies.-complaint history review: the complaint databases show there have been other events for the reported lot.Similar events have occurred for the catalog number and product family.These events were determined to be associated with ra 2012-067.Conclusions voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported abnormal ions level is considered to be under the scope of this recall.No further investigation is required.
 
Event Description
Plaintiff was implanted with a left rejuvenate modular hip stem.It¿s further alleged that blood work revealed elevated levels of cobalt and chromium.The patient¿s left hip was revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LRG TAP PRI MOD NCK 0DEG 30MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick, NJ NA
EI   NA
2018315000
MDR Report Key14165351
MDR Text Key289733187
Report Number0002249697-2022-00562
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2016
Device Catalogue NumberNLS-300000B
Device Lot Number35286402
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2090-2012
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-