Model Number 1000096 |
Device Problems
Filling Problem (1233); Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received for evaluation; however, device evaluation is not yet complete.A supplemental report will be submitted upon completion of the evaluation.
|
|
Event Description
|
It was reported that a minimum fill notification occurred after the user filled the cartridge with 280 units of insulin during the load sequence.Customer¿s blood glucose level was 328 mg/dl.Reportedly, the customer reverted to manual injections for insulin therapy.
|
|
Manufacturer Narrative
|
The failure investigation has been completed.Based on the analysis, the alleged issue could not be verified; however, a different issue was identified (damaged usb pins).
|
|