• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problem Wireless Communication Problem (3283)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/06/2022
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.During processing of this complaint, attempts were made to obtain patient weight, and but it was not received.
 
Event Description
It was reported the patient underwent a non-related surgical procedure wherein the ipg was not placed into surgery mode prior to the procedure.As a result, the patient's ipg was unable to communicate with external devices and the patient lost therapy.A manufacturer representative confirmed the issue and the ipg was deemed inoperable.Surgical intervention may occur at a later date to address the issue.
 
Event Description
Additional information reported surgical intervention occurred wherein the entire system was explanted to address the issue.
 
Manufacturer Narrative
Correction to h10: conclusion statement was omitted from final report in error.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14184802
MDR Text Key289853110
Report Number3006705815-2022-13665
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/19/2020
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000064236
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS ANCHOR (X2); SCS LEAD (X2)
Patient Outcome(s) Other;
Patient SexFemale
-
-