Model Number 24716
Device Problems Deflation Problem (1149); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2022
Event Type  malfunction  
Event Description
It was reported that balloon deflation failure occurred. The 95% stenosed target lesion was located in the mildly tortuous and mildly calcified anterior tibial artery. A 3. 0x220x90 (4f) sterling balloon catheter was advanced for dilatation. During the procedure, the balloon was inflated at 6 atmospheres for 1 minute; however, the balloon would not deflate. Tried double negative with the 3way and a 30cc attached as well. Tried a stiffer wire and eventually, was able to move the balloon enough to deflate some. The device was removed partially inflated from the patient and a kink was noticed on the device. The procedure was completed with another of same device. There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTERLING
Type of DeviceCATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14187111
MDR Text Key289884258
Report Number2134265-2022-04070
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 04/22/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number24716
Device Catalogue Number24716
Device Lot Number0028887465
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No