C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refw4255 showed no other similar product complaint(s) from this lot number.
|
|
Event Description
|
It was reported picc line without length markings, it was only missing on one piece.Detected in connection with insertion when it was to be cut, the part that lacked markings was cut off and saved but the picc line catheter could be used.
|
|
Event Description
|
It was reported picc line without length markings, it was only missing on one piece.Detected in connection with insertion when it was to be cut, the part that lacked markings was cut off and saved but the picc line catheter could be used.
|
|
Manufacturer Narrative
|
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of missing depth markings is confirmed and appears to be manufacturing related.One powerpicc segment and one product sticker label were returned for evaluation.The product sticker label indicates lot: (b)(4).The distal segment of the catheter returned appeared to have been trimmed where the 35 cm depth marker should be located; however, a section of the depth markers appeared to be missing.No depth markers were present proximal to the 42 cm depth marker.Microscopic observation of the missing depth markings regions did not reveal any evidence of ink transfer.The missing depth markers likely occurred during the manufacturing process; therefore, the complaint is confirmed.Bd is working closely with the manufacturing facility to prevent recurrence of the reported event.
|
|