Model Number N/A
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2022
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refw4255 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported picc line without length markings, it was only missing on one piece.Detected in connection with insertion when it was to be cut, the part that lacked markings was cut off and saved but the picc line catheter could be used.
 
Event Description
It was reported picc line without length markings, it was only missing on one piece.Detected in connection with insertion when it was to be cut, the part that lacked markings was cut off and saved but the picc line catheter could be used.
 
Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of missing depth markings is confirmed and appears to be manufacturing related.One powerpicc segment and one product sticker label were returned for evaluation.The product sticker label indicates lot: (b)(4).The distal segment of the catheter returned appeared to have been trimmed where the 35 cm depth marker should be located; however, a section of the depth markers appeared to be missing.No depth markers were present proximal to the 42 cm depth marker.Microscopic observation of the missing depth markings regions did not reveal any evidence of ink transfer.The missing depth markers likely occurred during the manufacturing process; therefore, the complaint is confirmed.Bd is working closely with the manufacturing facility to prevent recurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14190287
MDR Text Key289977153
Report Number3006260740-2022-01454
Device Sequence Number1
Product Code LJS
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K091324
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number2194108
Device Lot NumberREFW4255
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-