MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Failure to Disconnect (2541); Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/23/2020
Event Type  Injury  
Event Description
Information was received from a consumer and a healthcare provider (hcp) via a company representative regarding a patient receiving gablofen (2000 mcg/ml at161 mcg/day) via an implanted pump. The patient¿s medical history was intractable spasticity. The indication for pump use was intractable spasticity. The patient¿s mother reported that the patient had intermittent therapy that started about 10 days after the last pump replacement procedure. There were no reported environmental, external, or patient factors that may have led or contributed to the issue. The pump and catheter were replaced. It was noted that the sutureless catheter connector was damaged during explant. The issue was noted to be resolved, and it was indicated that the hcp had no further information to provide regarding the event.
 
Manufacturer Narrative
Concomitant medical products: product id: 8709sc, serial#: (b)(4), implanted: (b)(6) 2008, explanted: (b)(6) 2022, product type: catheter. Other relevant device(s) are: product id: 8709sc, serial/lot #: (b)(4), ubd: 06-may-2010, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14191765
MDR Text Key289958698
Report Number3004209178-2022-05161
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/14/2021
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Date Manufacturer Received04/19/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured06/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/25/2022 Patient Sequence Number: 1
-
-