PHILIPS ULTRASOUND EPIQ 7C; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  malfunction  
Event Description
A customer reported an x8-2t transducer would not articulate on an epiq 7c ultrasound system during use.A new transducer has been ordered to replace the suspect transducer to resolve this issue.There was no patient or user harm associated with this event.
 
Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow-up report upon its return and investigation completion.
 
Manufacturer Narrative
Evaluation of the transducer could not confirm the articulation issue as described by the customer.However, the device was found to be damaged.Visual inspection noted, paint chips on connector, scratches on mid-handle, protein build-up on i- tube, scratches on i-tube and scratches on tip shell.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ 7C
Type of Device
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key14194114
MDR Text Key289995819
Report Number3019216-2022-00048
Device Sequence Number1
Product Code IYN
UDI-Device Identifier00884838097872
UDI-Public00884838097872
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132304
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM
Device Catalogue Number795201
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-