Model Number 0998-00-0800-53
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2022
Event Type  malfunction  
Event Description
It was reported that during a daily check, the blood pressure (bp) cable for the cardiosave intra-aortic balloon pump (iabp) was unable to connect into the console. There was no patient involvement, and no adverse event was reported.
 
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate this unit. The fse replicated the reported issue and was unable to connect the bp adapter cable to the console. The bp connector had a tab blocking the cable from fully seating into the console. The fse noted that despite the issue, the iabp unit was still functional. Additionally, the fse observed external damage. The fse replaced the front end module and updated the software. The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer. A supplemental report will be submitted upon completion of our investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID, TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14197404
MDR Text Key290699917
Report Number2249723-2022-00929
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/25/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-