It was reported that the customer experienced an elevated blood glucose (bg) level; cause was not known.Customer's continuous glucose monitor read "high," however, specific bg value was not provided.A correction bolus via the pump was delivered to address bg.Additionally, it was reported that an o-ring was on the pneumatic tap.The customer was able to remove the o-ring from the pump and successfully reload the existing cartridge and resumed insulin therapy.It was also reported that insulin drips were not observed to be exiting the infusion set tubing during the load fill tubing process.Reportedly, the cartridge was changed to address the issue.Customer's blood glucose level was 303 mg/dl.
|