CARDIOVASCULAR SYSTEMS, INC. STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; CATHETER, PERIPHERAL, ATHERECTOMY
|
Back to Search Results |
|
Model Number PRD-SC30-MICRO |
Device Problem
Material Separation (1562)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 03/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The oad was returned to csi for analysis.Visual examination confirmed the reported driveshaft fracture.The location of the fracture could not be determined due to the severe stretching of the driveshaft.There was no other damage observed with the returned oad.Scanning electron microscopy analysis of the fractured filars shows evidence of ductile torsion which is an indication that the driveshaft was pulled to fracture.Therefore, the driveshaft fractured is considered to be the result of removal attempts by the user.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Patient age: 70.Age is approximate.(b)(4).
|
|
Event Description
|
A stealth 360 gen 2 peripheral orbital atherectomy device (oad) was used for treatment of a lesion in the anterior tibial (at) artery.Access was obtained via a contralateral femoral approach.The vessel was 100% occluded from the proximal femoral artery to the ankle joint, approximately 300mm.Fluoroscopy revealed moderate calcium.The lesion was primary wired with a non-csi wire.Following wire exchange, three atherectomy passes were performed in the proximal segment on low and medium speeds.The oad was advanced distally to the mid portion of the at and treatment was performed on low speed.The crown of the oad became entrapped in the vessel.Unsuccessful attempts were made to retrieve the oad, including glideassist and backward movement.The oad was forcefully removed, and the tip of the oad driveshaft fractured in the proximal at on the proximal end of the crown.The fragment was abandoned in vivo and not stented as the vessel was occluded.Angiography revealed no other vessels were compromised due to the fragment.Following the procedure, the patient was stable and was discharged.
|
|
Search Alerts/Recalls
|
|
|