The device was returned for investigation.Upon evaluation of the device, no image and black screen were noted.In addition, bending section cover leak, distal end crack, bending section cover cut, eyepiece body center off, and hole in the insertion tube were observed.The investigation is ongoing.Therefore; the root cause of the reported event cannot be determined at this time.However; if additional information becomes available, this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event was most likely caused by irradiation of the laser in the forceps channel and the resin in the ig serpentine melted.The event can be detected/prevented by following the instructions for use which state: 1) " inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities".2) "perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk".3) "do not irradiate the laser without viewing the tip of the laser probe and the aiming beam in the endoscopic image.This may cause patient injury, and the instrument channel may be damaged".Olympus will continue to monitor field performance for this device.
|