Model Number 383536 |
Device Problem
Retraction Problem (1536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2022 |
Event Type
malfunction
|
Event Description
|
It was reported 2 bd nexiva¿ closed iv catheter systems had difficult needle disengagement.The following information was provided by the initial reporter: "from what i understand the needle is not allowing the plastic cannula to release.".
|
|
Manufacturer Narrative
|
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter phone#: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported 2 bd nexiva¿ closed iv catheter systems had difficult needle disengagement.The following information was provided by the initial reporter: "from what i understand the needle is not allowing the plastic cannula to release.".
|
|
Manufacturer Narrative
|
H.6.Investigation: since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.A device history record review showed no non-conformances associated with this issue during the production of this batch.The complaint could not be confirmed and the root cause is undetermined.H3 other text : see h.10.
|
|
Search Alerts/Recalls
|