Model Number CLK-4
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  malfunction  
Event Description
A user facility reported to olympus that the halogen light source was not working.The issue was found by a technician during routine inspections.No patient involvement or injury was reported due to the event.
 
Manufacturer Narrative
The suspect device has been returned to olympus for evaluation and the investigation is in process.The olympus service center confirmed the customer's event and found the halogen lamp is not turning on.It was also noted that the air pump is not working due to a defective thermal fuse, the lamp base movement is not secure and the lamp socket, lamp stay and sticker are deformed.A definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that a faulty thermal fuse caused the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14238643
MDR Text Key299382307
Report Number8010047-2022-07192
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024511
UDI-Public04953170024511
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Was the Report Sent to FDA? No
Date Manufacturer Received06/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-