Model Number CLK-4 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Event Description
|
A user facility reported to olympus that the halogen light source was not working.The issue was found by a technician during routine inspections.No patient involvement or injury was reported due to the event.
|
|
Manufacturer Narrative
|
The suspect device has been returned to olympus for evaluation and the investigation is in process.The olympus service center confirmed the customer's event and found the halogen lamp is not turning on.It was also noted that the air pump is not working due to a defective thermal fuse, the lamp base movement is not secure and the lamp socket, lamp stay and sticker are deformed.A definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that a faulty thermal fuse caused the event.Olympus will continue to monitor field performance for this device.
|
|