• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL X-SMART PROTAPER INTROC.EUR HANDPIECE, DIRECT DRIVE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL X-SMART PROTAPER INTROC.EUR HANDPIECE, DIRECT DRIVE, AC-POWERED Back to Search Results
Catalog Number A100800000200
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2022
Event Type  malfunction  
Event Description
In this event it is reported that x-smart protaper introc. Eur won't hold files. Failure found before use on patient. No injury occurred.
 
Manufacturer Narrative
After inspection of device, head assembly, cartridge, head cap and bearing retainer replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameX-SMART PROTAPER INTROC.EUR
Type of DeviceHANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ CH-1338
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key14238802
MDR Text Key290556755
Report Number8031010-2022-00523
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 04/28/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA100800000200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/20/2022
Date Manufacturer Received04/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage

-
-