DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) shut down while the unit was plugged in.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
The full name of the event site in block e1 was shortened due to field character limit; the full name is providence (b)(6) center.Additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to evaluate the iabp.The fse observed that the battery did not run to factory specifications and replaced the lithium ion batteries to correct the issue.Subsequently, the fse performed all functional and safety checks to meet factory specifications and all testing passed.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.Patient height: 172 cm.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h4, h6, h10.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14239263
MDR Text Key290467344
Report Number2249723-2022-00955
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
IAB CATHETER
Patient Age68 YR
Patient SexMale
Patient Weight81 KG
-
-