DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) shut down while the unit was plugged in.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
The full name of the event site in block e1 was shortened due to field character limit; the full name is providence (b)(6) center.Additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to evaluate the iabp.The fse observed that the battery did not run to factory specifications and replaced the lithium ion batteries to correct the issue.Subsequently, the fse performed all functional and safety checks to meet factory specifications and all testing passed.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.Patient height: 172 cm.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, g7, h2, h4, h6, h10.
|
|
Event Description
|
N/a.
|
|